CVE-ID

CVE-2018-14781

• CVSS Severity Rating • Fix Information • Vulnerable Software Versions • SCAP Mappings • CPE Information
Description
Medtronic MiniMed MMT devices when paired with a remote controller and having the “easy bolus” and “remote bolus” options enabled (non-default), are vulnerable to a capture-replay attack. An attacker can capture the wireless transmissions between the remote controller and the pump and replay them to cause an insulin (bolus) delivery.
References
Note: References are provided for the convenience of the reader to help distinguish between vulnerabilities. The list is not intended to be complete.
Assigning CNA
ICS-CERT
Date Record Created
20180801 Disclaimer: The record creation date may reflect when the CVE ID was allocated or reserved, and does not necessarily indicate when this vulnerability was discovered, shared with the affected vendor, publicly disclosed, or updated in CVE.
Phase (Legacy)
Assigned (20180801)
Votes (Legacy)
Comments (Legacy)
Proposed (Legacy)
N/A
This is an record on the CVE List, which provides common identifiers for publicly known cybersecurity vulnerabilities.