CVE-ID

CVE-2017-6018

• CVSS Severity Rating • Fix Information • Vulnerable Software Versions • SCAP Mappings • CPE Information
Description
An open redirect issue was discovered in B. Braun Medical SpaceCom module, which is integrated into the SpaceStation docking station: SpaceStation with SpaceCom module (integrated as part number 8713142U), software versions prior to Version 012U000040, and SpaceStation (part number 8713140U) with installed SpaceCom module (part number 8713160U), software versions prior to Version 012U000040. The web server of the affected product accepts untrusted input which could allow attackers to redirect the request to an unintended URL contained within untrusted input.
References
Note: References are provided for the convenience of the reader to help distinguish between vulnerabilities. The list is not intended to be complete.
Assigning CNA
ICS-CERT
Date Record Created
20170216 Disclaimer: The record creation date may reflect when the CVE ID was allocated or reserved, and does not necessarily indicate when this vulnerability was discovered, shared with the affected vendor, publicly disclosed, or updated in CVE.
Phase (Legacy)
Assigned (20170216)
Votes (Legacy)
Comments (Legacy)
Proposed (Legacy)
N/A
This is an record on the CVE List, which provides common identifiers for publicly known cybersecurity vulnerabilities.