CVE-ID |
CVE-2017-14022
|
• CVSS Severity Rating • Fix Information • Vulnerable Software Versions • SCAP Mappings • CPE Information
|
Description |
An Improper Input Validation issue was discovered in Rockwell Automation FactoryTalk Alarms and Events, Version 2.90 and earlier. An unauthenticated attacker with remote access to a network with FactoryTalk Alarms and Events can send a specially crafted set of packets packet to Port 403/TCP (the history archiver service), causing the service to either stall or terminate.
|
References |
Note: References are provided for the convenience of the reader to help distinguish between vulnerabilities. The list is not intended to be complete.
|
|
Assigning CNA |
ICS-CERT |
Date Record Created |
20170830 |
Disclaimer: The record creation date may reflect when
the CVE ID was allocated or reserved, and does not
necessarily indicate when this vulnerability was
discovered, shared with the affected vendor, publicly
disclosed, or updated in CVE.
|
Phase (Legacy) |
Assigned (20170830) |
Votes (Legacy) |
|
Comments (Legacy) |
|
Proposed (Legacy) |
N/A |
This is an record on the CVE List, which provides common identifiers for publicly known cybersecurity vulnerabilities. |
|
For More Information: CVE Request Web Form (select "Other" from dropdown)
|