|
|
CVE-ID | ||
---|---|---|
CVE-2019-18241 |
• CVSS Severity Rating • Fix Information • Vulnerable Software Versions • SCAP Mappings • CPE Information
|
|
Description | ||
In Philips IntelliBridge EC40 and EC80, IntelliBridge EC40 Hub all versions, and IntelliBridge EC80 Hub all versions, the SSH server running on the affected products is configured to allow weak ciphers. This could enable an unauthorized attacker with access to the network to capture and replay the session and gain unauthorized access to the EC40/80 hub. | ||
References | ||
Note: References are provided for the convenience of the reader to help distinguish between vulnerabilities. The list is not intended to be complete. | ||
Assigning CNA | ||
ICS-CERT | ||
Date Record Created | ||
20191022 | Disclaimer: The record creation date may reflect when the CVE ID was allocated or reserved, and does not necessarily indicate when this vulnerability was discovered, shared with the affected vendor, publicly disclosed, or updated in CVE. | |
Phase (Legacy) | ||
Assigned (20191022) | ||
Votes (Legacy) | ||
Comments (Legacy) | ||
Proposed (Legacy) | ||
N/A | ||
This is an record on the CVE List, which provides common identifiers for publicly known cybersecurity vulnerabilities. | ||
You can also search by reference using the CVE Reference Maps.
|
||
For More Information: CVE Request Web Form (select "Other" from dropdown) |