CVE-ID |
CVE-2021-38393
|
• CVSS Severity Rating • Fix Information • Vulnerable Software Versions • SCAP Mappings • CPE Information
|
Description |
A Blind SQL injection vulnerability exists in the /DataHandler/HandlerAlarmGroup.ashx endpoint of Delta Electronics DIAEnergie Version 1.7.5 and prior. The application does not properly validate the user-controlled value supplied through the parameter agid before using it as part of an SQL query. A remote, unauthenticated attacker can exploit this issue to execute arbitrary code in the context of NT SERVICE\MSSQLSERVER.
|
References |
Note: References are provided for the convenience of the reader to help distinguish between vulnerabilities. The list is not intended to be complete.
|
|
Assigning CNA |
ICS-CERT |
Date Record Created |
20210810 |
Disclaimer: The record creation date may reflect when
the CVE ID was allocated or reserved, and does not
necessarily indicate when this vulnerability was
discovered, shared with the affected vendor, publicly
disclosed, or updated in CVE.
|
Phase (Legacy) |
Assigned (20210810) |
Votes (Legacy) |
|
Comments (Legacy) |
|
Proposed (Legacy) |
N/A |
This is a record on the CVE List, which provides common identifiers for publicly known cybersecurity vulnerabilities. |
|
For More Information: CVE Request Web Form (select "Other" from dropdown)
|