CVE-ID

CVE-2020-16224

• CVSS Severity Rating • Fix Information • Vulnerable Software Versions • SCAP Mappings • CPE Information
Description
In Patient Information Center iX (PICiX) Versions C.02, C.03, the software parses a formatted message or structure but does not handle or incorrectly handles a length field that is inconsistent with the actual length of the associated data, causing the application on the surveillance station to restart.
References
Note: References are provided for the convenience of the reader to help distinguish between vulnerabilities. The list is not intended to be complete.
Assigning CNA
ICS-CERT
Date Record Created
20200731 Disclaimer: The record creation date may reflect when the CVE ID was allocated or reserved, and does not necessarily indicate when this vulnerability was discovered, shared with the affected vendor, publicly disclosed, or updated in CVE.
Phase (Legacy)
Assigned (20200731)
Votes (Legacy)
Comments (Legacy)
Proposed (Legacy)
N/A