CVE-ID |
CVE-2018-18984
|
• CVSS Severity Rating • Fix Information • Vulnerable Software Versions • SCAP Mappings • CPE Information
|
Description |
Medtronic CareLink 2090 Programmer CareLink 9790 Programmer 29901 Encore Programmer, all versions, The affected products do not encrypt or do not sufficiently encrypt the following sensitive information while at rest PII and PHI.
|
References |
Note: References are provided for the convenience of the reader to help distinguish between vulnerabilities. The list is not intended to be complete.
|
|
Assigning CNA |
ICS-CERT |
Date Record Created |
20181106 |
Disclaimer: The record creation date may reflect when
the CVE ID was allocated or reserved, and does not
necessarily indicate when this vulnerability was
discovered, shared with the affected vendor, publicly
disclosed, or updated in CVE.
|
Phase (Legacy) |
Assigned (20181106) |
Votes (Legacy) |
|
Comments (Legacy) |
|
Proposed (Legacy) |
N/A |
This is an record on the CVE List, which provides common identifiers for publicly known cybersecurity vulnerabilities. |
|
For More Information: CVE Request Web Form (select "Other" from dropdown)
|